1.Heat up times are stated from ambient 22°C. 2. Timer settings are user pre-set (variable 0 to 999 minutes).
3. Timed cycle activated automatically once set point temperature has been reached.
4. Heating well dimensions 140w x 50d x 72h mm.
5. All aluminium blocks have a 6.5mm hole for a thermometer or probe.
6. Single Block systems, DBK-0 and DBK-10 to DBK-50, block dimensions 138w x 74d x 45h mm.
7. Double Block systems, DBK-0.2 to DBK-2.0, block dimensions 69w x 74d x 45h mm.
8. Double Block system, any two blocks can be fitted in the heating well.
Temperature Performance Notes
A. To verify the performance of the temperature control system the unit undergoes a factory calibration procedure which calculates the temperature values over the operating range of the equipment from two reference points.
B. A printed calibration record is available as an accessory option and can be ordered when unit is purchased quoting Stock No. CAL.
C. This record provides a measurement of precision about set point with maximum and minimum sensitivity readings obtained. Precision is defined as "(1) The ability of a measurement to be consistently reproduced" and "(2) The number of significant digits to which a value has been reliably measured."
D. This record does not provide or assures any form of accuracy. Accuracy is defined as "the ability of a measurement to match the actual value of the quantity being measured".
E. Where Accuracy is required use a calibrated external reference thermometer or measuring device and adjust set point to suit.
|