โพสเมื่อ
ชื่อสินค้า
Flame Photometer
รายละเอียด:
Linearisation: | |
Included in the software, and available on any of the 5 elements. Achievable on both single and Multi Point calibrations. | |
Aspiration Rate: | |
3-5.5ml/min | |
Optimal Range: | |
Single Point Calibration: | |
| |
Multi Point Calibration: | |
Na | |
0.05 | |
- 60ppm | |
| |
Na | |
0.05 | |
- 1000ppm | |
K | |
0.05 | |
- 100ppm | |
| |
K | |
0.05 | |
- 1000ppm | |
Li | |
0.1 | |
- 50ppm | |
| |
Li | |
0.1 | |
- 1000ppm | |
Ca | |
2.5 | |
- 100ppm | |
| |
Ca | |
2.5 | |
- 1000ppm | |
| |
| |
Ba | |
30 | |
- 3000ppm | | | |
Reproducibility: | |
<1% coefficient of variability for 20 consecutive samples over 10 minutes (after instrument stabilisation) at concentrations of 100ppm or less. | |
Limits of Detection: | |
Na | |
| |
0.02ppm. | |
K | |
| |
0.02ppm. | |
Li | |
| |
0.05ppm. | |
Ca | |
| |
1.0ppm. | |
Ba | |
| |
10ppm. | | | |
Time to Stability: | |
Less than 15 seconds after sample is introduced into the flame. | |
Drift: | |
Less than 1% per 30 minutes after instrument stabilisation. | |
Specifity: | |
Na/K/Li = <0.5% to each other when equal in concentration at <100ppm. | |
Number of Parameters Measured: | |
Simultaneous display of Na, K, Ca, Li, Ba. | |
Size: | |
51 cm high x 38 cm wide x 41 cm deep. | |
Weight: | |
Instrument = 14 kg Shipping = 21 kg | |
คำค้นหา:
photometer flame