BACHARACH ,เครื่องวัดไนตรัสออกไซด์ ,BACHARACH,Instruments and Controls/Laboratory Equipment
รหัสสินค้า
NANA-154365
ชื่อสินค้า
BACHARACH
แบรนด์
BACHARACH
โมเดล
Halogen Safe Entry Monitor
ผลิตที่
USA
รายละเอียด:
Halogen Safe Entry Monitor
Product Information
  •  Data Sheet
  •  Instruction Manual
  • Get Adobe® Reader®
  • Feedback Form
  • Warranty Registration

The Bacharach Halogen Safe Entry Monitor (hereafter referred to as the H-SEM unit) is a portable refrigerant gas monitor with a self contained purge-air supply. The H-SEM unit is designed to test an area for a specific refrigerant gas for the purpose of determining the presence and level of that gas in the area being tested.

The self contained purge-air supply allows an area to be tested for up to 40 minutes without needing to take the monitor to an area of fresh air for zeroing purposes. The monitor self-zeros every 4 minutes to ensure accurate gas measurements.

During normal operation the H-SEM unit displays the type of gas being monitored, along with displaying both the current gas level and the peak gas level detected in that area on its front panel LCD. The monitor retains a log of previous readings that can be easily accessed at a later time for analysis.

Front panel indicators and an audible alarm are provided to signal gas alarm and instrument fault conditions. The audible alarm can temporarily be silenced by a push of a button, but will reactivate if the gas level is still above the alarm trip point after a user defined time period has elapsed.

The H-SEM unit requires only minor periodic maintenance such as the occasional replacement of filters. The monitor incorporates active diagnostics that continuously check the system for proper operation. A front panel indicator is provided to alert an operator of system malfunctions, and fault codes are generated that enable the operator to identify the cause of the fault.

Features
  • Detects and measures CFC, HCFC, HFC and halogen gases
  • Measures all gases up to 1,000 ppm with a sensitivity of 1 ppm
  • Automatically logs up to 200 gas readings that can be later be recalled for analysis
  • Eliminates false alarms with use of non-dispersive IR source and sample draw system
  • Visual and audible gas alarm indictors that are turned ON when the detected gas level exceeds a user defined trip-point
  • Extensive self diagnostics, providing both visual and audible indications when a fault occurs
  • Battery powered, providing from 8?12 hours of operation
Specifications
  • Product Type: Portable refrigerant gas monitor with built-in purge-air source
  • Gas Library: R-11, R-12, R-21, R-22, R-23, R-113, R-114, R-123, R-124, R-134A, R-227, R-236fa, R-401A, R-402A, R-402B, R-404A, R-407A, R-407C, R-408A, R-409A, R-410A, R-500, R-502, R-503, R-507, R-508B, H-1211, H-1301, H-2402, N-1230, FA-188, CUSTOM
  • Measuring Range: All gases 0 to 1,000 ppm
  • Warm-Up Time: 15 minutes
  • Operating Time: 40 minutes (approx.) before the purge-air bag needs refilling
  • Detector Type: Infrared, Non-Dispersive
  • Sensitivity: 1 ppm
  • Accuracy: ±1 ppm
  • Response Time: 90% of response within 5 seconds; 100% in 7 seconds
  • Temperature Drift: 1.5 ppm per °C between purge cycles
  • System Noise: Less than 40dB(A) at 10 ft (3 m)
  • Operating Temp: 32 to 122°F (0 to 50°C)
  • Ambient Humidity: 5 to 90% RH (non-condensing)
  • Altitude Limit: 6,562 ft (2,000 m)
  • Power: DC power pack, provides 8?12 hours of operation
  • Power Consumption: 15 Watts
  • Front Panel: 3 Indicator lights:
    • ON ? Green LED flashes during warm-up, and then glows steady during normal operation
    • FAULT ? Yellow LED flashes when there is a system fault
    • ALARM ? Red LED flashes when the gas level is above its alarm setting
  • Audible Alarm: Internal audible alarm programmable for any of the following conditions:
    • OFF
    • FAULT/ALARM
    • ALARM
  • Dimensions: 8D x 19L x 5W inches (203 x 483 x 127 mm)
  • Weight: Less than 9 lbs (4 kg) including battery
  • Valid Calibration Period to Specifications: 12 months
  • Warranty: 2 Years from date of shipment
Ordering Information
Part NumberDescription
3015-4484 Halogen Safe Entry Monitor
3015-3420 External Filter
07-1654 Internal Particulate Filter
3015-4654 Battery Pack & Charger
3015-4405 Battery Charge Cable
3015-4474 Soft Carrying Case w/ Shoulder Strap