รายละเอียด:
Instrument Type : T8DS
Optical System : Double beam
Wavelength Range : 190~900nm
Wavelength Accuracy : ± 0.3nm
Wavelength Reproducibility : ? 0.1nm
Spectral Bandwidth : 0.1, 0.2, 0.5, 1.0, 2.0, 5.0nm
Photometric Mode Transmittance, Absorbance, Energy Concentration, All Using UVWin Software
Photometric Range -4.0~4.0Abs
Photometric Accuracy ±0.002A (0~0.5A), ±0.004A (0.5~1.0A), ±0.3%T (0~100%T)
Photometric Reproducibility ±0.001A (0~0.5A)
Photometric Noise ±0.0004A (500nm) 30min warm-up
Baseline Flatness ±0.001A (200~850nm)
Baseline Stability ±0.0008A/h (500nm, 0Abs), 2hr warm-up
Stray light ? 0.01%T (220nm Nal, 340nm NaNO2)
Standard Functionality No stand alone function
Cell Holder Fixed position sample and reference
Detector Photo multiplier tube
Light Source Tungsten Halogen and Deuterium arc lamps
Display No display
Printer Not available
PC Interface : RS232
Software Support UV-Win
Power Supply Switchable 120~230VAC 50~60Hz
Weight : 43Kg
Dimensions (Width, Depth, Height) : 545mm, 580mm, 270mm
ภาพประกอบ: