Applications | |
Metering of liquids and gases, viscous and corrosive media.
Applications: chemical, petrochemical, oil and gas industries
The TME Series is also used for precise dosing, loading and unloading systems as well as for custody transfer applications.
|
Reliability | |
Robust Design, immune to vibration effects and to pipeline generated stresses, rugged housing (cast iron) |
Special features | |
Simultaneous metering of mass flow, density, temperature and volumetric flow.
Inexpensive integrated heating jacket technology: steam, liquid, electrical
Removable sensor housing cover
|
Technical data | |
Sensor
End connections: Flanges acc. EN 1092, ASME B16.5, DIN2512
Nominal pressure: PN 40, ASME Cl 150 / 300 / 600 lbs
Process temperature: -40°C to + 180°C (-40°F to +356°F)
Ambient temperature (junction box): -40°C to + 100°C (-40°F to +212°F)
Ingress protection: IP 65 (EN60529), (NEMA 4X )
Wetted parts: 1.4404 (316 L) / 1.4571 (316 Ti)
Secondary containment: cast iron
CE-mark: Pressure Equipment Directive 97/23/EG
Ranges: 60 kg/h up to 60.000 kg/h (liquid)
Transmitter
Direct or remote mounted
Power supply 19 - 36 VDC, 24 VAC +/- 20%,90 - 265 VAC
Outputs: galvanically isolated
analog: 2 x 0/4-20 mA
Binary 1: active, potential free 24 V=, max. 200 mA
Passive: optocoupler Ui=30 V, Ii=200mA, Pi=3 W
Frequency: 1 KHz
Binary 2: passive, optocoupler Ui=30 V, Ii=200mA, Pi=3 W
Status: passive, optocoupler Ui=30 V, Ii=200mA, Pi=3 W
Input binary: counter reset
Ambient temperature: -20°C bis +60°C/+80°C (-4°F to +140°F/176°F)
Ingress protection: IP 68 (EN60529) (NEMA6)
Accuracy: Liquid: ± 0,15%/ ± 0,05% of reading ± zero point stability
Gas: ± 0,5% of reading ± zero point stability
Density (liquid): ± 0,005 g/cm³/ ± 0,003 g/cm³
Volume: ± 0,2% of reading ± zero point stability
|
Approvals and certificates | |
Explosion protection sensor:
circuits intrinsically safe
DMT 01 ATEX E 149 X
II 1/2G EEx ia IIC T6-T2
(Approval for Zone 0 inside flow tubes)
CE-mark: Pressure Equipment Directive 97/23/EG
Explosion protection transmitter:
BVS 05 ATEX E 021 X
Increased safety: EEx e (connection): II (1)2G EEx de [ia] IIC/IIB T6-T3
Explosion proof: EEx d (connection): II (1)2G EEx d [ia] IIC/IIB T6-T3
Signal output / input: intrinsically safe or not intrinsically safe
FM XP-AIS / I / 1 / A B C D / T* : CD 06100
FMC XP-AIS / I / 1 / C D / T* : CD 06101
NEPSI Approval Cert No. GYJ06477
CE-mark: Explosion Protection Directive 94/9/EG
EMC-Directive 89/336/EWG
Material certificates 3.1, 3.2 acc. DIN EN10204:2004
Custody transfer approval
|
Communication | |
HART®
Profibus-PA
Modbus RTU (RS 485)
|